Infant Care Equipments H-2000 Baby Warmers Infant Incubator
Product Feature
- Servo Control Mode: air temperature, skin temperature;
- Parameter Displaying: set temperature, air temperature, skin temperature, internal humidity, heat power rate;
- Audible & Visual Alarms: power failure, sensor failure, higher & lower temp, air & skin temperature override, fan failure, system failure;
- >37ºC Preset Mode: power failure reset memory, alarm silence, keypad lock, self-test function;
- Hood with dual slope; 6 hand-ports;
- RS-232 Output Connector (use for network system);
- IV poles;
- Protective guard rail;
- Storage cabinets
Technical Specification
Power requirement | ≤400VA |
Temperature control mode | Air and skin servo control |
Air Temp Control Range | 20.0ºC-37.0ºC |
Temperature override mode to | 37.1ºC-39.0ºC |
Skin Temp Control Range | 20.0ºC-37.0ºC |
Temperature override mode to | 37.1ºC-38.0ºC |
Skin temp sensor accuracy | ≤0.3ºC |
Temperature fluctuation | ≤0.5ºC |
Mattress temperature uniformity | ≤0.8ºC |
Temperature rise time | ≤45 min |
Noise level within hood environment | ≤55dB |
Alarms | Power failure alarm, Fan failure alarm, Sensor failure alarm |
High air temperature alarm | 3.0ºC |
Low air temperature alarm | -3.0ºC |
High skin temperature alarm | 1.0ºC |
Low skin temperature alarm | -1.0ºC |
Over air temperature alarm | ≤38.0ºC; ≤40.0ºC |
Over skin temperature alarm | ≤38.0ºC; ≤39.0ºC |
System failure alarm | |
Alarm silence function |
Specifications of hood
- Mattress to Hood: 450 mm
- Observation side of hood: Two incline
- Operate port: 6 (2 Iris ports)
- Tubing grommets: 4
- Front access panel: 826×310 mm
- Mattress Tray Size: 630×360 mm
- Infant bed tilt adjusting ±10° step-less
Operational condition
- Environment temperature: 20ºC- 30ºC
- Environment relative humidity: 30%-75%
- Environment air velocity of flow: <0.3m/s
Others
- Mattress tray Max. load 5kg
- Transfusion pole Max. load 2kg
- Cabinet Max. load 6kg
Others similar products
Incubator
Incubator
৳ 210,000.00
৳ 325,000.00
৳ 1,048,000.00
Techno Health – Head office
House-42, Lake Drive Road
Sector-07, Uttara, Dhaka-1230
Cell No: 01842756014
Facebook: technohealth.com.bd
Techno Health (Topkhana Road Branch)
Bagdad Surgical Market
21/A, (1st Floor) Mahbub Plaza
Topkhana Road, Dhaka-1000
01812754847
Reviews
There are no reviews yet.